• Home
  • Introduction
  • Software
  • About me

Looking At Nothing

A SA(X)S Weblog

Category:

  • Clear Presentation Initiative
  • LookingAtNothing Weblog
  • Matlab
  • misc.
  • Movies
  • Paper review
  • Papers
  • Particle size distribution
  • Posters
  • Pushbutton machines
  • sampleholder design
  • SAXSGUI software
  • software
  • Uncategorized

Archives:

  • August 2010
  • July 2010
  • June 2010
  • April 2010
  • March 2010
  • February 2010
  • January 2010
  • December 2009
  • November 2009
  • October 2009
  • September 2009
  • August 2009
  • February 2009
  • November 2008
  • October 2008
  • July 2008
  • May 2008
  • February 2008
  • December 2007
  • November 2007
  • October 2007
  • September 2007
  • August 2007
  • July 2007
  • June 2007
  • May 2007
  • April 2007

Blogroll:

  • Nico Adams
  • Presentation zen

Meta:

  • RSS
  • Comments RSS
  • Valid XHTML
  • XFN

Impossible angles in the Porod relationship, an open question

August 28th, 2007 by Brian

The unbiquitous porod relationship, as published in many places, is often written as:
lim I = K/s^4.

Here, s is the scattering vector, described by s=2/lambda sin(theta) with lambda the radiation wavelength, and theta as half the scattering angle.

Strangely, though, the limit in the porod relationship is often written as the limit when s becomes infinity. Looking at the description of s, however, one sees immediately that this is not possible, whilst the maximum value of s (with a constant wavelength) is 2/lambda for a scattering angle of 90 degrees.

So how am I to read that porod relationship then? as a limit where s goes to 2/lambda, or where s goes to an unreachable infinity?

Addendum:
Theoretically, the scattering vector $q$ or $s$ can indeed go to infinity, but only if the radiation wavelength approaches zero. Nevertheless, it remains interesting why the limit is not to 2/lambda, but instead is written as it goes to infinity.

Posted in LookingAtNothing Weblog | 1 Comment

Untapped potential

August 20th, 2007 by Brian

I think this weblog style communication has a lot of untapped potential. I am seriously considering the following:

– Opening a publicly accessible weblog on SAXS, where anyone can post anything he deems interesting for the rest of the readership to read/see/know. This would be significantly different from a forum in that it is not intended for too much discussion afterwards.
– Opening a private weblog in which I can discuss topics with my supervisors regarding my Ph.D. project. This way we can have communication outside of the standard fora, such as meetings and co.

What are your opinions on this? Leave a comment!

Posted in LookingAtNothing Weblog | 1 Comment

Pushbutton machines – aligning pinholes

August 16th, 2007 by Brian

Many consumers (dare I say Bio-people here?) would love to have a pushbutton SAXS machine. At the moment, however, there are a few issues here that have to be taken care of. One of them is the rather cumbersome pinhole alignment procedure.

After replacing a source, realigning the monochromator crystal, or after changing of the geometry of a SAXS machine, the collimation pinholes need to be realigned. Thus, the part of the beam with the highest intensity is to be selected for this. The standard method is as illustrated in this figure:

Standard pinhole alignment procedure

The uncollimated beam (A) produces a rather large spot on the detector. By moving in the first pinhole, most of the surrounding radiation is cut off (B). Then the second pinhole is moved into position (C), which cuts the beam into the desired shape. The parasitic scattering generated by the second pinhole is then removed by introducing the third pinhole (D) resulting in a collimated system.

There are a few issues with this method:
– If the pinhole positions are not well known, introducing pinholes two and three might become ever more tricky. This is because the only way to know where the center of the second pinhole is, is when the pinhole opening actually crosses the beam that has been reduced in size by the previous pinhole. This can get very difficult in the larger geometries, where a pinhole of say .3 mm in diameter has to be moved over a range of millimeters in two directions in order to find the beam.
– You usually find out that you moved the beam too low (for example due to misalignment of the monochromator mirror), only when you are trying to move pinhole three into position. At that point, when you can no longer move the pinhole further to center it around the beam, you have to disassemble the collimation system and realign the mirror, undoing all of the collimation work that you have done before.
– If the pinholes are aligned by the countrate over the detector, a low countrate when aligning the second pinhole may in the end cause a large error in the placement of the second pinhole. This can later, when the entire pinhole collimation system is in place, no longer be corrected (i.e. when you find out that the intensity of the collimated beam is not as much as it was once upon a time with a different collimation).

I think this could be improved using the following method:
Pinhole individual alignment procedure

Here, the position of each pinhole with respect to the beam center and the detector is subsequently determined (A through D). Then, a region with high beam intensity on the detector can be selected, and the pinholes can be moved there, using a spherical coordinate system to move the pinholes with the source point as center. This way, an alternative position of the beam can be selected simply by moving all the pinholes in this coordinate system to cut out the new part of the beam (F) without having to realign the pinholes.

Thus, you can quickly find out if the beam is misaligned (a region can be indicated on the detector output display that all the pinholes can reach), and an alternative part of the beam can be selected. The drawback of this method is that it would require a little more than a simple counter on the detector, i.e. it would require image processing algorithms to find circle-shaped shadows of the pinholes cutting the primary beam.

More specifically, the image processing would need to correlate the center of those pinhole shadows on the detector grid, to the motor positions of the pinhole alignment motors. Furthermore, the length of the sections between the pinholes is to be measured and input in order to be able to move the pinholes around an arbitrary center (e.g. the source point) in a spherical coordinate system.

I think, however, that this method would allow for a much more robust push-button approach to the alignment of pinholes. I hope this would make realisation of a complete pushbutton system a step closer to reality. I will attempt to implement this at the Risø system in the near future.

Leave your comments at the bottom!

Posted in LookingAtNothing Weblog, Pushbutton machines | 2 Comments

New Site

August 11th, 2007 by Brian

I have moved the contents of the iWeb designed website to a WordPress one. Whilst Wordpress does not look as fancy as the iWeb stuff did, it now does allow you, the viewers, to actively contribute by leaving comments!

In addition to that, posting has become a little more convenient and perhaps I can add a publicly accessible forum on a later date.

Leave your comments!

Posted in LookingAtNothing Weblog | 1 Comment

Opinion on a recent paper by Takeshi Morita

August 2nd, 2007 by Brian

This paper, Morita et al., j. app. cryst., 40, 791–795, 2007 shows a short note on a system capable of measuring the absorption factor of a sample. They measure the attenuated beam by means of a beamstop with a photodiode in it, and the primary beam intensity by means of an ionisation chamber situated before the sample. Unfortunately, the beamstop design seems rather bulky, and would therefore obstruct approximately 1/8th of the complete scattering pattern.

I am reminded at this point of a system designed by Hermans, Heikens and Weidinger c.a. 1959 (Hermans et al., j. pol. sci., 35, 145–165, 1959). They created an apparatus (slit-collimated if I’m not mistaken) that contained no less than three photographic films. The first film would measure the primary beam intensity as it scattered through a thin polymethyl methacrylate film (as a sort of primitive ionisation chamber). The attenuated beam measurement was performed by measuring the backscattered radiation from the beamstop (a tungsten wire). The third film was used to measure the scattering pattern itself.

This allowed them to do a similar determination of all required parameters for a measurement on absolute scale. It is nice to see that a reiteration of this technique has appeared, the newer detectors allowing for less tedious calibrations.

I feel that not enough people measure on an absolute scale at the moment, thereby forfeiting information on their samples that might have proven useful. Hopefully such efforts as described by Morita et al. will initiate more initiatives to allow users to painlessly and effortlessly do measurements on an absolute scale.

Posted in LookingAtNothing Weblog | 1 Comment

 